RGO - Revista Gaúcha de Odontologia2021
Application of computed tomography dose index as dosimetric parameter in dental tomography
Rodrigo Antonio Pereira MAURO e Alessandro Martins da COSTA
em inglês
resumo
ABSTRACT Objective Quantify the CTDI from a reference beam, correcting this value for several beam widths from the ratio of CTDI100 values measured in the air, and verifying the underestimation of the dose comparing CTDI100 and CTDI300, applied to dental CBCT. Methods i-CAT and Prexion 3D tomographs, 100mm pencil ionization chamber, electrometer. Firstly for beam above 40 mm, CTDIW,NT from CTDIW,Ref, multiplied by the ratio of CTDIAR measurements to N.T widths and reference, was estimated. In second, CTDI100 and CTDI300 are obtained by displacing the ionization chamber along the beam with spacing intervals equal to 100 mm to cover sufficiently large integration intervals for CBCT protocols, and a comparison is made through the modified efficiency. Results CTDIAR,100,Ref averaged 53% higher than CTDIW,Ref, due to attenuation of the beam by the simulator object, and the ratio between CTDIAR,100,Ref and CTDIW,Ref is greater than 1, being this constant relationship and validating the method for dosimetry in quality control tests. For the second method, CTDI100 greatly underestimates the dose deposited on the central axis, where CTDI300 covers all beam sizes and stands out in relation to CTDI100 to more accurately quantify the radiation levels emitted. Conclusions The IEC method is applicable to quality control, facilitating practice, and optimizing time and resources. CTDI300 is a better dose descriptor than CTDI100, and should be implemented for CBCT modalities when used.
como citar · ABNT NBR 6023
MAURO, Rodrigo Antonio Pereira; COSTA, Alessandro Martins da. Application of computed tomography dose index as dosimetric parameter in dental tomography. RGO - Revista Gaúcha de Odontologia, 2021. DOI: 10.1590/1981-86372021000820180112.
Referências
26 obras citadas
- 1DJ Brenner e EJ Hall. Computed tomography – An increasing source of radiation exposure. N Engl J Med, 2007quem mais cita →
- 2R Smith-Bindman, J Lipson, R Marcus e mais 3. Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med, 2009quem mais cita →
- 3DJ Brenner, I Shuryak e AJ Einstein. Impact of reduced patient life expectancy on potential cancer risks from radiologic imaging. Med Phys, 2011quem mais cita →
- 4MS Pearce, JA Salotti, MP Little e mais 3. Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: A retrospective cohort study. Lancet, 2012quem mais cita →
- 5TB Shope, RM Gagne e GC Johnson. A method for describing the doses delivered by transmission x-ray compute tomography. Med Phys, 1981quem mais cita →
- 6MF McNitt-Gray. AAPM/RSNA physics tutorial for residents: Topics in CT. RadioGraphics, 2002quem mais cita →
- 7Commission IE. Medical electrical equipment-Part 2-44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography. 2009quem mais cita →
- 8RL Dixon, JA Anderson, DM Bakalyar e mais 3. Comprehensive methodology for the evaluation of radiation dose in x-ray computed tomography. Rep AAPM Task Gr, 2010quem mais cita →
- 9ID McLean. Status of computed tomography dosimetry for wide cone beam scanners. International Atomic Energy Agency, 2011quem mais cita →
- 10RL Dixon. Restructuring CT dosimetry-A realistic strategy for the future Requiem for the pencil chamber. Med Phys, 2006quem mais cita →
- 11DJ Brenner, CH McCollough e CG Orton. It is time to retire the computed tomography dose index (CTDI) for CT quality assurance and dose optimization. Med Phys, 2006quem mais cita →
- 12S Mori, M Endo, K Nishizawa e mais 3. Enlarged longitudinal dose profiles in cone-beam CT and the need for modified dosimetry. Med Phys, 2005quem mais cita →
- 13M Loubele, R Jacobs, F Maes e mais 3. Image quality vs radiation dose of four cone beam computed tomography scanners. Dentomaxillofacial Radiol, 2008quem mais cita →
- 14E Helmrot, A Thilander-klang, H Sciences e mais 1. Methods for Monitoring patient dose. Radiat Prot Dosimetry, 2010quem mais cita →
- 15S Han, B Lee, G Shin e mais 3. Dose area product measurement for diagnostic reference levels and analysis of patient dose in dental radiography. Radiat Prot Dosimetry, 2012quem mais cita →
- 16R Pauwels, J Beinsberger, B Collaert e mais 3. Effective dose range for dental cone beam computed tomography scanners. Eur J Radiol, 2012quem mais cita →
- 17WO Batista, MVT Navarro e AF Maia. Development of a phantom and a methodology for evaluation of depth kerma and kerma index for dental cone beam computed tomography. Radiat Prot Dosimetry, 2013quem mais cita →
- 18M Vinicius, L Oliveira, W Otto e mais 2. Air kerma area product in cone beam computed tomography. 2014quem mais cita →
- 19JB Ludlow e C Walker. Assessment of phantom dosimetry and image quality of i-CAT FLX cone-beam computed tomography. Am J Orthod Dentofac Orthop, 2013quem mais cita →
- 20W Leitz, B Axelsson e G Szendrö. Computed tomography dose assessment-a practical approach. Radiat Prot Dosimetry, 1995quem mais cita →
Quem cita esta obra
nenhum artigo do acervo cita esta obra — ainda
É sua obra?
Receba um aviso quando um artigo novo citar este trabalho.
também serve para acompanhar o autor que você estuda