ABSTRACT Introduction: The harmonization of equipment is recommended in clinical laboratory practice aiming for the homogeneity of results when similar or equivalent analyzers are used to perform routine testing. Objectives: To conduct a study of equivalence between the biochemical analyzers Labmax 240® (E1) and Labmax 240 Premium® (E2) through the matching results and the statistical value analysis of dosages. Materials and methods: We evaluated tests with glucose, total cholesterol, triglycerides, uric acid, aspartate transaminase (AST), alanine transaminase (ALT) and lactate dehydrogenase (LDH), all with 40 repeated measurements, performed in both equipments. The Clinical and Laboratory Standards Institute (CLSI) EP09-3A protocol was used to conduct the comparison test between E1 and E2 equipment, with subsequent evaluation of the results for statistical analysis determining the Pearson correlation coefficient (r) and indexes comparison error with EP Evaluator® software. Results: Regarding the values of the Pearson correlation coefficient, all tests showed a strong correlation between equipment with r > 0.989, except for the dosage of LDH (r = 0.982). This dosage failed not because the value of r, but due to the values obtained in the error index being larger than the total errors index allowed. Discussion: Compared to clinical criteria, the results of the analyzers are approximately equal, but this control process must be done continuously in order to prevent and track random errors within the laboratory routine. Conclusion: The process of harmonization of multiple devices that perform the same laboratory parameters is essential for ensuring quality and reliability of laboratory results and should be standardized and included in routine clinical analysis laboratories.

como citar · ABNT NBR 6023

SCAPIN, Luisa D. et al. Harmonization study of results between biochemical analyzers Labmax 240® and Labmax 240 Premium®. Jornal Brasileiro de Patologia e Medicina Laboratorial, 2018. DOI: 10.5935/1676-2444.20180028.

Referências

13 obras citadas · 3 no acervo

  1. 1J Westgard e T Darcy. The truth about quality: medical usefulness and analytical reliability of laboratory tests. Clin Chim Acta., 2004quem mais cita →
  2. 2FA Berlitz. Controle da qualidade no laboratório clínico: alinhando melhoria de processos, confiabilidade e segurança do paciente. J Bras Patol Med Lab., 2010no acervo →
  3. 3S Vilela. Equipamentos automatizados - resultados de exames laboratoriais mais precisosquem mais cita →
  4. 4CA Oliveira e ME Mendes. Equivalência de sistema analítico. Gestão da fase analítica do laboratório: como assegurar a qualidade na prática, 2011quem mais cita →
  5. 5Clinical and Laboratory Standards Institute. Method comparison and bias estimation using patient samplesquem mais cita →
  6. 6C Ricos, V Alvarez e F Cava. Current databases on biologic variation: pros, cons and progress. Scand J Clin Lab Invest., 1999quem mais cita →
  7. 7KF Vieira, ES Shitara, ME Mendes e mais 1. A utilidade dos indicadores da qualidade no gerenciamento de laboratórios clínicos. J Bras Patol Med Lab, 2011no acervo →
  8. 8C Weber. Garantia da qualidade no setor de bioquímica do laboratório de análises clínicas de um hospital público de Porto Alegre. 2012quem mais cita →
  9. 9Agência Nacional de Vigilância Sanitária. RDC 302, de 13 de outubro de 2005quem mais cita →
  10. 10Sociedade Brasileira de Patologia Clínica e Medicina Laboratorial. Normativas e avaliação PALCquem mais cita →
  11. 11Sociedade Brasileira de Análises Clínicas. Acreditação do sistema de qualidade do laboratório clínicoquem mais cita →
  12. 12HJJ Lopes. Garantia e controle da qualidade no laboratório clínicoquem mais cita →
  13. 13GA Campana e CP Oplustil. Conceitos de automação na medicina laboratorial: revisão de literatura. J Bras Patol Med Lab, 2011no acervo →

Quem cita esta obra

nenhum artigo do acervo cita esta obra — ainda

É sua obra?

Receba um aviso quando um artigo novo citar este trabalho.

também serve para acompanhar o autor que você estuda
Acompanhar as citações