International Journal of Cardiovascular Sciences2024
Agreement Between a Portable Cholesterol Device and Laboratory-Based Testing in Older Adults
Odilon Abrahin, Naicha Stefanie Félix Souza, Rejane P. Abrahin e Alex Harley Crisp
em inglês
resumo
Abstract Background: It is crucial to monitor blood lipid levels accurately in older adults to assess cardiovascular risk. Although several portable strip-based devices are commercially available, their accuracy has not been well established. Objective: To evaluate the agreement of the Mission Cholesterol device and standard laboratory method results, using samples from older adults. Methods: Forty-nine patients (42 females) with an average age of 70 ± 8 years were included. The participants were instructed to fast for 12 hours before undergoing venous and capillary blood sampling for lipid analysis (total cholesterol [TC], high-density lipoprotein cholesterol [HDL-C], low-density lipoprotein cholesterol [LDL-C], and triglyceride [TG] levels). The agreement between the Mission Cholesterol device and laboratory test results was assessed using the concordance correlation coefficient (CCC) and Bland–Altman plots. The accuracy levels were assessed based on the percentage bias recommended by the National Cholesterol Education Program (NCEP). Results: The TC measurement had a better accuracy level (bias = 10.91%; CCC = 0.89) than other lipid measurements. However, none of them met the NCEP's acceptable standards for bias and concordance (CCC ≥ 0.90). Furthermore, the device demonstrated modest overall agreement in classifying lipid risk according to clinical reference values (agreement rates: TC, 67.3%; HDL-C, 65.3%; LDL-C, 49.0%; and TG, 61.2%). Conclusion: The Mission Cholesterol device exhibited insufficient agreement levels for effective lipid profile monitoring and screening in older adults.
como citar · ABNT NBR 6023
ABRAHIN, Odilon et al. Agreement Between a Portable Cholesterol Device and Laboratory-Based Testing in Older Adults. International Journal of Cardiovascular Sciences, 2024. DOI: 10.36660/IJCS.20240101.
Referências
17 obras citadas · 1 no acervo
- 1SM Grundy, NJ Stone, AL Bailey e mais 3. 2018 AHA/ACC/AACVPR/AAPA/ABC/ACPM/ADA/AGS/APhA/ASPC/NLA/PCNA Guideline on the Management of Blood Cholesterol: Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines. J Am Coll Cardiol, 2019quem mais cita →
- 2J Zheng, J Wang, Y Zhang e mais 3. The Global Burden of Diseases Attributed to High Low-density Lipoprotein Cholesterol from 1990 to 2019.. Front Public Health, 2022quem mais cita →
- 3ACMGN Sá, CS Gomes, AD Moreira e mais 2. Prevalence and Factors Associated with Self-reported Diagnosis of High Cholesterol in the Brazilian Adult Population: National Health Survey 2019. Epidemiol Serv Saude, 2022no acervo →
- 4M Mudau, A Genis, A Lochner e mais 1. Endothelial Dysfunction: The Early Predictor of Atherosclerosis. Cardiovasc J Afr, 2012quem mais cita →
- 5GL Myers, MM Kimberly, PP Waymack e mais 3. A Reference Method Laboratory Network for Cholesterol: A Model for Standardization and Improvement of Clinical Laboratory Measurements. Clin Chem, 2000quem mais cita →
- 6S Kurstjens, E Gemen, S Walk e mais 3. Performance of Commercially-available Cholesterol Self-tests. Ann Clin Biochem, 2021quem mais cita →
- 7JJZ Liao. Sample Size Calculation for an Agreement Study. Pharm Stat, 2010quem mais cita →
- 8WT Friedewald, RI Levy e DS. Fredrickson. Estimation of the Concentration of Low-density Lipoprotein Cholesterol in Plasma, without use of the Preparative Ultracentrifuge. Clin Chem, 1972quem mais cita →
- 9National Cholesterol Education Program. Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel III). National I, 2002quem mais cita →
- 10LI Lin. A Concordance Correlation Coefficient to Evaluate Reproducibility. Biometrics, 1989quem mais cita →
- 11JM Bland e DG. Altman. Statistical Methods for Assessing Agreement between Two Methods of Clinical Measurement. Lancet, 1986quem mais cita →
- 12AR. Caldwell. SimplyAgree: An R Package and Jamovi Module for Simplifying Agreement and Reliability Analyses. J Open Source Softw, 2022quem mais cita →
- 13H. Wickham. Ggplot2. Wires Comput Stat, 2011quem mais cita →
- 14P Quartey, D Gator, MA Mpiani e mais 1. Evaluation of the Analytical Performance of the Mission Cholesterol Meter for Serum Lipids Using NCEP Criteria. Int J Med Health Res, 2020quem mais cita →
- 15IR Kupke, B Kather e S. Zeugner. On the Composition of Capillary and Venous Blood Serum. Clin Chim Acta, 1981quem mais cita →
- 16IR Kupke, S Zeugner, A Gottschalk e mais 1. Differences in Lipid and Lipoprotein Concentrations of Capillary and Venous Blood Samples. Clin Chim Acta, 1979quem mais cita →
- 17M Hyvärinen, HK Juppi, S Taskinen e mais 3. Metabolic Health, Menopause, and Physical Activity-a 4-year Follow-up Study. Int J Obes, 2022quem mais cita →
Quem cita esta obra
nenhum artigo do acervo cita esta obra — ainda
É sua obra?
Receba um aviso quando um artigo novo citar este trabalho.
também serve para acompanhar o autor que você estuda