Research on Biomedical Engineering2018
Modeling and FPGA-based implementation of an efficient and simple envelope detector using a Hilbert Transform FIR filter for ultrasound imaging applications
Amauri Amorin Assef, Breno Mendes Ferreira, Joaquim Miguel Maia e Eduardo Tavares Costa
comunicação breve · em inglês
resumo
Abstract Introduction Although the envelope detection is a widely used method in medical ultrasound (US) imaging to demodulate the amplitude of the received echo signal before any back-end processing, novel hardware-based approaches have been proposed for reducing its computational cost and complexity. In this paper, we present the modeling and FPGA implementation of an efficient envelope detector based on a Hilbert Transform (HT) approximation for US imaging applications. Method The proposed model exploits both the symmetry and the alternating zero-valued coefficients of a HT finite impulse response (FIR) filter to generate the in-phase and quadrature components that are necessary for the envelope computation. The hardware design was synthesized for a Stratix IV FPGA, by using the Simulink and the integrated DSP Builder toolbox, and implemented on a Terasic DE4-230 board. The accuracy of our algorithm was evaluated by the normalized root mean square error (NRMSE) cost function in comparison with the conventional method based on the absolute value of the discrete-time analytic signal via FFT. Results An excellent agreement was achieved between the theoretical simulations with the experimental result. The NRMSE was 0.42% and the overall FPGA utilization was less than 1.5%. Additionally, the proposed envelope detector is capable of generating envelope data at every FPGA clock cycle after 19 (0.48 µs) cycles of latency. Conclusion The presented results corroborate the simplicity, flexibility and efficiency of our model for generating US envelope data in real-time, while reducing the hardware cost by up to 75%.
como citar · ABNT NBR 6023
ASSEF, Amauri Amorin et al. Modeling and FPGA-based implementation of an efficient and simple envelope detector using a Hilbert Transform FIR filter for ultrasound imaging applications. Research on Biomedical Engineering, 2018. DOI: 10.1590/2446-4740.02417.
Referências
14 obras citadas
- 1AA Assef, JM Maia e ET Costa. Initial experiments of a 128-channel FPGA and PC-based ultrasound imaging system for teaching and research activities. 2016quem mais cita →
- 2AA Assef, JM Maia, FK Schneider e mais 2. Design of a 128-channel FPGA-based ultrasound imaging beamformer for research activities. 2012quem mais cita →
- 3JH Chang, JT Yen e KK Shung. A novel envelope detector for high-frame rate, high-frequency ultrasound imaging. IEEE Trans Ultrason Ferroelect Freq Control., 2007quem mais cita →
- 4LS DeBrunner e Y Wang. Optimizing filter order and coefficient length in the design of high performance FIR filters for high throughput FPGA implementations. 2006quem mais cita →
- 5V Hans. Signal processing of complex modulated ultrasonic signals. Fluid mechanics of flow metering, 2005quem mais cita →
- 6MA Hassan e YM Kadah. Digital signal processing methodologies for conventional digital medical ultrasound imaging system. Am J Biomed Eng., 2013quem mais cita →
- 7JA Jensen, O Holm, LJ Jerisen e mais 10. Ultrasound research scanner for real-time synthetic aperture data acquisition. IEEE Trans Ultrason Ferroelect Freq Control., 2005quem mais cita →
- 8P Levesque e M Sawan. Real-time hand-held ultrasound medical-imaging device based on a new digital quadrature demodulation processor. IEEE Trans Ultrason Ferroelectr Freq Control., 2009quem mais cita →
- 9L Marple. Computing the discrete-time "analytic" signal via FFT. IEEE Trans Sig Process., 1999quem mais cita →
- 10J McClellan, TW Parks e L Rabiner. A computer program for designing optimum FIR linear phase digital filters. IEEE Trans Audio Electroacoust., 1973quem mais cita →
- 11W Qiu, Y Yu, FK Tsang e mais 1. An FPGA-based open platform for ultrasound biomicroscopy. IEEE Trans Ultrason Ferroelectr Freq Control., 2012quem mais cita →
- 12M Schlaikjer, JP Bagge, OM Sorensen e mais 1. Trade off study on different envelope detectors for B-mode imaging. 2003quem mais cita →
- 13MA Soderstrand, LG Johnson, H Arichanthiran e mais 2. Reducing hardware requirement in FIR filter design. 2000quem mais cita →
- 14H Zhou e YF Zheng. An efficient quadrature demodulator for medical ultrasound imaging. Front Inf Technol Electr Eng., 2015quem mais cita →
Quem cita esta obra
nenhum artigo do acervo cita esta obra — ainda
É sua obra?
Receba um aviso quando um artigo novo citar este trabalho.
também serve para acompanhar o autor que você estuda