Brazilian Journal of Radiation Sciences2025

A Monte Carlo Simulation study on the design and optimization of collimators for non-collinear cascade gamma-ray correlation emissions in medical imaging

L. L. Nkuba e I. J. Lugendo

em inglês

cita 36 obras

resumo

Abstract This study focused on designing and optimizing collimators for cascade gamma-ray imaging through Monte Carlo simulations. The trapezoidal-shaped collimator blocks, designed in the Geant4 application for emission tomography (GATE) environment, were attached to a simulated small animal GATE - PET model. The collimators were optimized by simulating septa thicknesses from 0.2 mm to 1.2 mm, in 0.2 mm increments. A 1.0 MBq 111In source having radius of 0.25 mm was used as the cascade gamma-ray emitter. Sixteen trapezoidal tungsten collimator blocks were designed, each with a 16.31 mm × 37.5 mm surface facing the detector crystals, and a 12.33 mm × 37.5 mm surface facing the scanned object. Each block featured 105 parallel rectangular holes arranged in a 7 × 15 array, with a length of 10.0 mm, resulting in a ring-like collimator with a 41.0 mm outer radius. The designed collimator, intended for small animal imaging, prioritizes resolution. Hence, a collimator with 1.0 mm septa and hole sizes of 1.5 mm × 0.7 mm, offering spatial resolutions of 7.6 mm and 4.1 mm in the axial and transaxial directions, respectively, was chosen. The collimators demonstrated energy resolution of approximately 8.96% and 10.10% at 171.3 keV and 245.4 keV, respectively, within a 10% energy resolution threshold set during simulations. Besides, the reconstructed source positions ranged from 81.1% to 100% of the true simulated source positions within the field of view. The optimized collimator design presents a viable solution for imaging small animals’ internal organs, with sizes exceeding 7.6 mm.

como citar · ABNT NBR 6023

NKUBA, L. L.; LUGENDO, I. J. A Monte Carlo Simulation study on the design and optimization of collimators for non-collinear cascade gamma-ray correlation emissions in medical imaging. Brazilian Journal of Radiation Sciences, 2025. DOI: 10.15392/2319-0612.2025.2573.

Referências

36 obras citadas

  1. 1R. B. Pahlka, S. C. Kappadath e O. R. Mawlawi. A Monte Carlo simulation of coincidence detection and imaging of gamma-ray cascades with a scintillation camera. Biomed. Phys. Eng. Express, 2018quem mais cita →
  2. 2J. R. Saffer, H. H. Barrett, H. B. Barber e mais 1. Surgical probe design for a coincidence imaging system without a collimator. Image Vis. Comput, 1992quem mais cita →
  3. 3T. Butz. Nuclear quadrupole interactions studied by time differential perturbed angular correlations of γ-rays. Zeitschrift fur Naturforsch. - Sect. A J. Phys. Sci, 1996quem mais cita →
  4. 4L. Hemmingsen, K. N. Sas e E. Danielsen. Biological Applications of Perturbed Angular Correlations of γ-Ray Spectroscopy. Chem. Rev, 2004quem mais cita →
  5. 5C. C. Chiang, C. C. Chuang, Y. C. Ni e mais 3. Time of flight dual photon emission computed tomography. Sci. Rep, 2020quem mais cita →
  6. 6M. Uenomachi, K. Shimazoe, K. Ogane e mais 1. Simultaneous multi-nuclide imaging via double-photon coincidence method with parallel hole collimators. Sci. Rep, 2021quem mais cita →
  7. 7C. Rangacharyulu, Thi Khanh L. Lai, K. Olshanoski e mais 8. Development of a Nuclear Gamma Cascade Correlations Based Medical Imaging System - A Novel Modality with Potentials for Replacement of SPECT and PET. International Conference on Clinical PET-CT and Molecular Imaging in the Era of Theranostics (IPET), Book of Abstract-IPT 2020, IAEA-CN-285/110, 2021quem mais cita →
  8. 8K. Shimazoe, M. Uenomachi, Y. Mizumachi e mais 5. Double Photon Emission Coincidence Imaging using GAGG-SiPM pixel detectors. J. Instrum, 2017quem mais cita →
  9. 9R. S. Santhosh, D. S. Shyam, T. Fukuchi e mais 3. Design and Optimization of a Collimator for a New PET system using GATE simulation. PROCEEDINGS, 64TH DAE BRNS SYMPOSIUM ON NUCLEAR PHYSICS, 2019 Lucknow (Uttar Pradesh), India, 2019quem mais cita →
  10. 10E. Brady e M. Deutsch. The angular correlation of successive gamma-rays. Phys. Rev, 1950quem mais cita →
  11. 11I. Schmitz-Feuerhake. Studies on three-dimensional scintigraphy with gamma - gamma - coincidences. Phys. Med. Biol, 1970quem mais cita →
  12. 12W. G. Monahan e M. D. Powell. Three-Dimensional Imaging of Radionuclide Distribution by Gamma-Gamma Coincidence Detection. 1972quem mais cita →
  13. 13S. Yamamoto, S. Okumura, T. Watabe e mais 7. Development of a prototype Open-close positron emission tomography system. Rev. Sci. Instrum, 2015quem mais cita →
  14. 14R. Massari, A. D'Elia, A. Soluri e mais 1. Super Spatial Resolution (SSR) method for small animal SPECT imaging: A Monte Carlo study. Nucl. Instruments Methods Phys. Res. Sect. A Accel. Spectrometers, Detect. Assoc. Equip, 2020quem mais cita →
  15. 15F. Scopinaro, R. Massari, A. D. Varvarigou e mais 3. High-resolution small animal single photon emission computed tomography: uptake of [99mTc] bombesin and [123I] ioflupane by rat brain. Q J Nucl Med Mol Imaging, 2007quem mais cita →
  16. 16K. Van Audenhaege, R. Van Holen, S. Vandenberghe e mais 3. Review of SPECT collimator selection, optimization, and fabrication for clinical and preclinical imaging. Med Phys, 2015quem mais cita →
  17. 17L. L. Nkuba, K. Olshanoski, T. Fukuchi e mais 4. Monte Carlo simulations of the non-collinear gamma-ray cascade emissions for medical imaging. 11th INTERNATIONAL CONFERENCE ON ISOTOPES (11-ICI) AND EXPO, Saskatoon, Canada, July 23-27, 2023, 2023quem mais cita →
  18. 18K. Olshanoski, L. L. Nkuba, N. P. Dang e mais 5. Collimator design for gamma-ray cascade angular correlations in medical imaging. J. Instrum, 2023quem mais cita →
  19. 19I. Buvat e I. Castiglioni. Monte Carlo simulations in SPET and PET. Q J Nucl Med, 2002quem mais cita →
  20. 20G. Santin, D. Strul, D. Lazaro e mais 3. GATE, a Geant4-based simulation platform for PET integrating movement and time management. IEEE Nucl. Sci. Symp. Med. Imaging Conf, 2003quem mais cita →
1–20 de 36 obras citadas

Quem cita esta obra

nenhum artigo do acervo cita esta obra — ainda

É sua obra?

Receba um aviso quando um artigo novo citar este trabalho.

também serve para acompanhar o autor que você estuda
Acompanhar as citações