Brazilian Dental Journal2015

An Indirect Method to Measure Abutment Screw Preload: A Pilot Study Based on Micro-CT Scanning

Carlos Eduardo E. Rezende, Jason Alan Griggs, Yuanyuan Duan, Amanda M. Mushashe, Gisele Maria Correr Nolasco, Ana Flávia Sanches Borges e José Henrique Rubo

em inglês

cita 20 obras

resumo

This study aimed to measure the preload in different implant platform geometries based on micro-CT images. External hexagon (EH) implants and Morse Tapered (MT) implants (n=5) were used for the preload measurement. The abutment screws were scanned in micro-CT to obtain their virtual models, which were used to record their initial length. The abutments were screwed on the implant with a 20 Ncm torque and the set composed by implant, abutment screw and abutment were taken to the micro-CT scanner to obtain virtual slices of the specimens. These slices allowed the measurement of screw lengths after torque application and based on the screw elongation. Preload values were calculated using the Hooke's Law. The preloads of both groups were compared by independent t-test. Removal torque of each specimen was recorded. To evaluate the accuracy of the micro-CT technique, three rods with known lengths were scanned and the length of their virtual model was measured and compared with the original length. One rod was scanned four times to evaluate the measuring method variation. There was no difference between groups for preload (EH = 461.6 N and MT = 477.4 N), but the EH group showed higher removal torque values (13.8±4.7 against 8.2±3.6 Ncm for MT group). The micro-CT technique showed a variability of 0.053% and repeatability showed an error of 0.23 to 0.28%. Within the limitations of this study, there was no difference between external hexagon and Morse taper for preload. The method using micro-CT may be considered for preload calculation.

como citar · ABNT NBR 6023

REZENDE, Carlos Eduardo E. et al. An Indirect Method to Measure Abutment Screw Preload: A Pilot Study Based on Micro-CT Scanning. Brazilian Dental Journal, 2015. DOI: 10.1590/0103-6440201300274.

Referências

20 obras citadas

  1. 1CJ Goodacre, G Bernal, K Rungcharassaeng e mais 1. Clinical complications with implants and implant prostheses. J Prosthet Dent, 2003quem mais cita →
  2. 2S Gracis, K Michalakis, P Vigolo e mais 3. Internal vs. external connections for abutments/reconstructions: a systematic review. Clin Oral Implants Res, 2012quem mais cita →
  3. 3MS. Schwarz. Mechanical complications of dental implants. Clin Oral Impl Res, 2000quem mais cita →
  4. 4PP. Binon. Implants and components: Entering the new millennium. Int J Oral Maxillofac Implants, 2000quem mais cita →
  5. 5RL Sakaguchi e SE. Borgersen. Nonlinear contact analysis of preload in dental implant screws. Int J Oral Maxillofac Implants, 1995quem mais cita →
  6. 6MR. Norton. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin Oral Implants Res, 1997quem mais cita →
  7. 7F Sutter, HP Weber, J Sorensen e mais 1. The new restorative concept of ITI dental implants system: Design and engineering. Int J Periodont Res Dent, 1997quem mais cita →
  8. 8MS Pita, RB Anchieta, VAR Barão e mais 3. Prosthetic platforms in implant dentistry. J Craniofac Surg, 2011quem mais cita →
  9. 9A. Khraisat. Stability of the implant/abutment joint in a single-tooth external-hexagon implant system: Clinical and mechanical review. Clin Implant Dent Relat Res, 2004quem mais cita →
  10. 10JE Haack, RL Sakaguchi, T Sun e mais 1. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants, 1995quem mais cita →
  11. 11LC Breeding, DL Dixon, EW Nelson e mais 1. Torque required to loosen single-tooth implant abutment screws before and after simulated function. Int J Prosthodont, 1993quem mais cita →
  12. 12EW. Colling. The physical metallurgy of titanium alloys. 1984quem mais cita →
  13. 13JP Geng, KB Tan e GR. Liu. Application of finite element analysis in implant dentistry: A review of the literature. J Prosthet Dent, 2001quem mais cita →
  14. 14AR Coppedê, AC Faria, MG Mattos e mais 3. Mechanical comparison of experimental conical-head abutment screws with conventional flat-head abutment screws for external-hex and internal tri-channel implant connections: An in vitro evaluation of loosening torque. Int J Oral Maxillofac Implants, 2013quem mais cita →
  15. 15EI Weiss, D Kozak e MD. Gross. Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent, 2000quem mais cita →
  16. 16LA Lang, B Kang, RF Wang e mais 1. Finite element analysis to determine implant preload. J Prosthet Dent, 2003quem mais cita →
  17. 17JRP Jorge, VAR Barão, JA Delben e mais 1. The role of implant/abutment system on torque maintenance of retention screws and vertical misfit of implant-supported crowns before and after mechanical cycling. Int J Oral Maxillofac Implants, 2013quem mais cita →
  18. 18A Cantwell e JA. Hobkirk. Preload loss in gold prosthesis-retaining screws as a function of time. Int J Oral Maxillofac Implants, 2004quem mais cita →
  19. 19SC Kano, P Paul Binon, G Bonfante e mais 1. Effect of casting procedures on screw loosening in UCLA-type abutments. J Prosthodont, 2006quem mais cita →
  20. 20JA Delben, EA Gomes, VAR Barão e mais 1. Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws. Int J Oral Maxillofac Implants, 2011quem mais cita →

Quem cita esta obra

nenhum artigo do acervo cita esta obra — ainda

É sua obra?

Receba um aviso quando um artigo novo citar este trabalho.

também serve para acompanhar o autor que você estuda
Acompanhar as citações