Química Nova2012
Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants
Ana Gabriela Reis Solano, Gisele Rodrigues da Silva, Silvia Ligório Fialho, Armando da Silva Cunha Júnior e Gérson Antônio Pianetti
em inglês
resumo
A method using HPLC-UV was developed and validated for the determination of etoposide incorporated into polycaprolactone implants. The method was carried out in isocratic mode using a C18 column (250 x 4.6 mm; 5 µm), at 25 ºC, with acetonitrile and acetic acid 4% (70:30) as mobile phase, a flow rate of 2 mL/min, and UV detection at 285 nm. The method was linear (r² > 0.99) over the range of 5 to 65 µg/mL, precise (RSD < 5%), accurate (recovery of 98.7%), robust, selective regarding excipient of the sample, and had a quantitation limit equal to 1.76 µg/mL. The validated method can be successfully employed for routine quality control analyses.
como citar · ABNT NBR 6023
SOLANO, Ana Gabriela Reis et al. Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants. Química Nova, 2012. DOI: 10.1590/S0100-40422012000600032.
Referências
6 obras citadas
- 9United States Pharmacopoeia. 2011quem mais cita →
- 10British pharmacopoeia. 2011quem mais cita →
- 12Brasil. Resolução RE nº899, 29/5/2003: Guia para validação de métodos analíticos e bioanalíticos. Diário Oficial da União, 2003quem mais cita →
- 16International Conference on Harmonisation. Topic Q2B validation of analytical procedures: methodology. 1996quem mais cita →
- 19G. W. Snedecor e W. G. Cochran. Statistical methods. 1996
- 20W. J. Youden e E. H. Steiner. Statistical manual of AOAC - Association of Official Analytical Chemistry. 1975quem mais cita →
Quem cita esta obra
nenhum artigo do acervo cita esta obra — ainda
É sua obra?
Receba um aviso quando um artigo novo citar este trabalho.
também serve para acompanhar o autor que você estuda