Química Nova2011

Development and validation of a high performance liquid chromatographic method for determination of cyclosporine-A from biodegradable intraocular implants

Juliana Barbosa Saliba, Armando da Silva Cunha Junior, Elionai Cassiana de Lima Gomes, Herman Sander Mansur e Gisele Rodrigues da Silva

em inglês

cita 3 obras

resumo

An HPLC method was developed and validated aiming to quantify the cyclosporine-A incorporated into intraocular implants, released from them; and in direct contact with the degradation products of PLGA. The separation was carried out in isocratic mode using acetonitrile/water (70:30) as mobile phase, a C18 column at 80 ºC and UV detection at 210 nm. The method provided selectivity based on resolution among peaks. It was linear over the range of 2.5-40.0 µg/mL. The quantitation and detection limits were 0.8 and 1.2 µg/mL, respectively. The recovery was 101.8% and intra-day and inter-day precision was close to 2%.

como citar · ABNT NBR 6023

SALIBA, Juliana Barbosa et al. Development and validation of a high performance liquid chromatographic method for determination of cyclosporine-A from biodegradable intraocular implants. Química Nova, 2011. DOI: 10.1590/S0100-40422011000100026.

Referências

3 obras citadas

  1. 27The United States Pharmacopeia. 2008quem mais cita →
  2. 28J. E. F Reynolda. Martindale the extra pharmacopoeia. 1989quem mais cita →
  3. 29Validation of Analytical Procedures: Text and Methodology. 2005quem mais cita →

Quem cita esta obra

nenhum artigo do acervo cita esta obra — ainda

É sua obra?

Receba um aviso quando um artigo novo citar este trabalho.

também serve para acompanhar o autor que você estuda
Acompanhar as citações