Química Nova2011
Development and validation of a high performance liquid chromatographic method for determination of cyclosporine-A from biodegradable intraocular implants
Juliana Barbosa Saliba, Armando da Silva Cunha Junior, Elionai Cassiana de Lima Gomes, Herman Sander Mansur e Gisele Rodrigues da Silva
em inglês
resumo
An HPLC method was developed and validated aiming to quantify the cyclosporine-A incorporated into intraocular implants, released from them; and in direct contact with the degradation products of PLGA. The separation was carried out in isocratic mode using acetonitrile/water (70:30) as mobile phase, a C18 column at 80 ºC and UV detection at 210 nm. The method provided selectivity based on resolution among peaks. It was linear over the range of 2.5-40.0 µg/mL. The quantitation and detection limits were 0.8 and 1.2 µg/mL, respectively. The recovery was 101.8% and intra-day and inter-day precision was close to 2%.
como citar · ABNT NBR 6023
SALIBA, Juliana Barbosa et al. Development and validation of a high performance liquid chromatographic method for determination of cyclosporine-A from biodegradable intraocular implants. Química Nova, 2011. DOI: 10.1590/S0100-40422011000100026.
Referências
3 obras citadas
Quem cita esta obra
nenhum artigo do acervo cita esta obra — ainda
É sua obra?
Receba um aviso quando um artigo novo citar este trabalho.
também serve para acompanhar o autor que você estuda