Revista Brasileira de Oftalmologia2026

Biometric accuracy in cataract surgery: a comparative study between the IOLMaster 500 and Argos

Nicoli Lopes de Oliveira, Camyla Lemos Budib, Juliana Carolina Contrera, Marcello Novoa Colombo-Barboza, Guilherme Novoa Colombo-Barboza e Bernardo Kaplan Moscovici

em inglês

cita 30 obras, 3 no acervo

resumo

ABSTRACT Purpose: To assess the agreement between IOLMaster 500 and Argos for ocular biometry parameters Methods: This retrospective study included 104 patients (205 eyes) who underwent biometry with both devices. Measurements included axial length (AL), keratometry (K1, K2), and ACD. Anterior chamber depth (ACD) measurements were compared directly as reported by each device, without additional correction for corneal thickness. Results: No significant differences were found for AL (23.49 ± 1.29 mm versus 23.63 ± 1.35 mm; p = 0.737), K1 (43.36 ± 1.68 D versus 43.41 ± 1.61 D; p = 0.769), or K2 (44.28 ± 1.60 D versus 44.17 ± 1.58 D; p = 0.575). However, ACD values differed significantly (2.98 ± 0.41 mm versus 3.18 ± 0.36 mm; p = 0.009). Conclusion: IOLMaster 500 and Argos yield comparable AL and keratometry values, supporting their interchangeable use for these parameters. However, due to significant differences and bias in ACD, this parameter should be interpreted with caution and should not be considered directly interchangeable between devices. These results highlight the importance of consistent biometry protocols in IOL power calculation.

como citar · ABNT NBR 6023

OLIVEIRA, Nicoli Lopes de et al. Biometric accuracy in cataract surgery: a comparative study between the IOLMaster 500 and Argos. Revista Brasileira de Oftalmologia, 2026. DOI: 10.37039/1982.8551.20260094.

Referências

30 obras citadas · 3 no acervo

  1. 1World Health Organization (WHO). Global data on visual impairments 2010. 2012quem mais cita →
  2. 2S Norrby. Sources of error in intraocular lens power calculation. J Cataract Refract Surg, 2008quem mais cita →
  3. 3RB Melles, JT Holladay e WJ Chang. Accuracy of intraocular lens calculation formulas. Ophthalmology, 2018quem mais cita →
  4. 4G Freeman e K Pesudovs. The impact of cataract severity on measurement acquisition with the IOLMaster. Acta Ophthalmol Scand, 2005quem mais cita →
  5. 5N Hirnschall, R Varsits, B Doeller e mais 1. Enhanced penetration for axial length measurement of eyes with dense cataracts using swept source optical coherence tomography: a consecutive observational study. Ophthalmol Ther, 2018quem mais cita →
  6. 6HJ Shammas e KJ Hoffer. Repeatability and reproducibility of biometry and keratometry measurements using a noncontact optical low-coherence reflectometer and keratometer. Am J Ophthalmol, 2012quem mais cita →
  7. 7M Porwolik, A Porwolik e E Mrukwa-Kominek. Evaluation of selected biometric parameters in cataract patients—a comparison between Argos® and IOLMaster 700®: two swept-source optical coherence tomography-based biometers. Medicina (Kaunas), 2024quem mais cita →
  8. 8N Sorkin, A Achiron, M Abumanhal e mais 3. Comparison of two new integrated SS-OCT tomography and biometry devices. J Cataract Refract Surg, 2022quem mais cita →
  9. 9A Galzignato, E Lupardi, KJ Hoffer e mais 3. Repeatability of new optical biometer and agreement with 2 validated optical biometers, all based on SS-OCT. J Cataract Refract Surg, 2023quem mais cita →
  10. 10J Románek e K Sluková. Comparison of optical biometers Argos and IOL Master 700. Cesk Slov Oftalmol, 2021quem mais cita →
  11. 11CM Yang, DH Lim, HJ Kim e mais 1. Comparison of two swept-source optical coherence tomography biometers and a partial coherence interferometer. PLoS One, 2019quem mais cita →
  12. 12X Liao, Y Peng, B Liu e mais 2. Agreement of ocular biometric measurements in young healthy eyes between IOLMaster 700 and OA-2000. Sci Rep, 2020quem mais cita →
  13. 13W Stopyra, A Langenbucher e A Grzybowski. Intraocular lens power calculation formulas—a systematic review. Ophthalmol Ther, 2023quem mais cita →
  14. 14AD Fișuș, ND Hirnschall e O Findl. Comparison of 2 swept-source optical coherence tomography-based biometry devices. J Cataract Refract Surg, 2021quem mais cita →
  15. 15DL Cooke e TL Cooke. Comparison of 9 intraocular lens power calculation formulas. J Cataract Refract Surg, 2016quem mais cita →
  16. 16CL Budib, CHP Oliani, GO Ferreira e mais 3. Reliability of a Scheimpflug-based tomographer compared with a Placido-based topographer and specular microscopy in healthy eyes. Arq Bras Oftalmol, 2026no acervo →
  17. 17N Shetty, L Kaweri, A Koshy e mais 3. Repeatability of biometry measured by three devices and its impact on predicted intraocular lens power. J Cataract Refract Surg, 2021quem mais cita →
  18. 18DG Altman e JM Bland. Measurement in medicine: the analysis of method comparison studies. Statistician, 1983quem mais cita →
  19. 19P Jaycock, RL Johnston, H Taylor e mais 3. The Cataract National Dataset electronic multi-centre audit of 55,567 operations: updating benchmark standards of care in the United Kingdom and internationally. Eye (Lond), 2009quem mais cita →
  20. 20World Medical Association. World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. JAMA, 2013quem mais cita →
1–20 de 30 obras citadas

Quem cita esta obra

nenhum artigo do acervo cita esta obra — ainda

É sua obra?

Receba um aviso quando um artigo novo citar este trabalho.

também serve para acompanhar o autor que você estuda
Acompanhar as citações