Research on Biomedical Engineering2017

Automated and observer based light field indicator edge evaluation in diagnostic X-ray equipment

Márcio Bottaro, Balázs Vince Nagy, Fernanda Cristina Salvador Soares e Danilo Cabral Rosendo

em inglês

cita 10 obras

resumo

Abstract Introduction To analyze edge detection and optical contrast calculation of light field-indicators used in X-ray via automated- and observer-based methods, and comparison with current standard approaches, which do not give exact definition for light field edge determination. Methods Automated light sensor array was used to measure the penumbra zone of the edge in the standard X-ray equipment, while trained and naïve human observers were asked to mark the light field edge according to their own determination. Different interpretations of the contrast were then calculated and compared. Results In contrast to automated measurements of edge definition and detection, measurements by human observers showed large inter-observer variation independent of their training with X-ray equipment. Different contrast calculations considering the different edge definitions gave very different contrast values. Conclusion As the main conclusion, we propose a more exact edge definition of the X-ray light field, corresponding well to the average human observer’s edge determination. The new edge definition method with automated systems would reduce human variability in edge determination. Such errors could potentially affect the approval of X-ray equipment, and also increase the radiation dose. The automated measurement based on human observers’ edge definition and the corresponding contrast calculation may lead to a more precise light field calibration, which enables reduced irradiation doses on radiology patients.

como citar · ABNT NBR 6023

BOTTARO, Márcio et al. Automated and observer based light field indicator edge evaluation in diagnostic X-ray equipment. Research on Biomedical Engineering, 2017. DOI: 10.1590/2446-4740.07116.

Referências

10 obras citadas

  1. 1MJ Balderson, DP Spencer, I Nygren e mais 1. Quality assurance using a photodiode array. Journal of Applied Clinical Medical Physics, 2011quem mais cita →
  2. 2H Brookfield, A Manning-Stanley e A England. Light beam diaphragm collimation errors and their effects on radiation dose for pelvic radiography. Radiologic Technology, 2015quem mais cita →
  3. 3M Davidson e JA Whiteside. Human brightness perception near sharp contours. Journal of the Optical Society of America, 1971quem mais cita →
  4. 4Food and Drug Administration. Code of Federal Regulations Title 21. Food and Drugs. Part 1020. Performance Standard for Ionizing Radiation Emitting Products. 2015quem mais cita →
  5. 5S Grossberg e E Mingolla. Neural dynamics of form perception: boundary completion, illusory figures, and neon color spreading. Psychological Review, 1985quem mais cita →
  6. 6International Electrotechnical Commission. IEC 60601-2-54:2009 - medical electrical equipment. particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy. 2009quem mais cita →
  7. 7K Kisielewicz, A Truszkiewicz, S Wach e mais 1. Evaluation of dose area product vs. patient dose in diagnostic X-ray units. Physica Medica, 2011quem mais cita →
  8. 8T Kron e S Ferguson. Where is the light field edge: perception of different operators on different surfaces. Medical Dosimetry, 2000quem mais cita →
  9. 9SP MacEvoy, WJ Kim e MA Paradiso. Integration of surface information in primary visual cortex. Nature Neuroscience, 1998quem mais cita →
  10. 10MK Woo e CS McParland. An optical edge detector for the accurate delineation of the light field in radiotherapy machines. Medical Physics, 2001quem mais cita →

Quem cita esta obra

nenhum artigo do acervo cita esta obra — ainda

É sua obra?

Receba um aviso quando um artigo novo citar este trabalho.

também serve para acompanhar o autor que você estuda
Acompanhar as citações