Parameter estimation of an artificial respiratory system under mechanical ventilation following a noisy regime
Marcus Henrique Victor Júnior, Luiz Alberto Forgiarini Junior, Toru Miyagi Kinjo, Marcelo Britto Passos Amato, Takashi Yoneyama e Harki Tanaka
em inglês
resumo
Abstract Introduction: This work concerns the assessment of a novel system for mechanical ventilation and a parameter estimation method in a bench test. The tested system was based on a commercial mechanical ventilator and a personal computer. A computational routine was developed do drive the mechanical ventilator and a parameter estimation method was utilized to estimate positive end-expiratory pressure, resistance and compliance of the artificial respiratory system. Methods The computational routine was responsible for establishing connections between devices and controlling them. Parameters such as tidal volume, respiratory rate and others can be set for standard and noisy ventilation regimes. Ventilation tests were performed directly varying parameters in the system. Readings from a calibrated measuring device were the basis for analysis. Adopting a first-order linear model, the parameters could be estimated and the outcomes statistically analysed. Results Data acquisition was effective in terms of sample frequency and low noise content. After filtering, cycle detection and estimation took place. Statistics of median, mean and standard deviation were calculated, showing consistent matching with adjusted values. Changes in positive end-expiratory pressure statistically imply changes in compliance, but not the opposite. Conclusion The developed system was satisfactory in terms of clinical parameters. Statistics exhibited consistent relations between adjusted and estimated values, besides precision of the measurements. The system is expected to be used in animals, with a view to better understand the benefits of noisy ventilation, by evaluating the estimated parameters and performing cross relations among blood gas, ultrasonography and electrical impedance tomography.
como citar · ABNT NBR 6023
VICTOR JÚNIOR, Marcus Henrique et al. Parameter estimation of an artificial respiratory system under mechanical ventilation following a noisy regime. Research on Biomedical Engineering, 2015. DOI: 10.1590/2446-4740.0581.
Referências
24 obras citadas
- 1A Albanese, N Karamolegkos, S Haider e mais 2. Real-time noninvasive estimation of intrapleural pressure in mechanically ventilated patients: a feasibility study. 2013quem mais cita →
- 2C Barr, M Çetinkaya-Rundel e D. Diez. OpenIntro statistics. 2012quem mais cita →
- 3A Beda, PM Spieth, T Handzsuj e mais 4. A novel adaptive control system for noisy pressure-controlled ventilation: a numerical simulation and bench test study. Intensive Care Medicine, 2010quem mais cita →
- 4AR Carvalho e WA Zin. Respiratory system dynamical mechanical properties: modeling in time and frequency domain. Biophysical Reviews, 2011quem mais cita →
- 5RL Cordioli, E Akoumianaki e L Brochard. Nonconventional ventilation techniques. Current Opinion in Critical Care, 2013quem mais cita →
- 6B Diong, H Nazeran, P Nava e mais 1. Modelling human respiratory impedance. IEEE Engineering in Medicine and Biology Magazine, 2007quem mais cita →
- 7J Fernández, D Miguelena, H Mulett e mais 2. Adaptive support ventilation: State of the art review. Indian Journal of Critical Care Medicine, 2013quem mais cita →
- 8M Gama de Abreu, PM Spieth, P Pelosi e mais 7. Noisy pressure support ventilation: A pilot study on a new assisted ventilation mode in experimental lung injury. Critical Care Medicine, 2008quem mais cita →
- 9G Girija, JR Raol e J Singh. Modelling and parameter estimation of dynamic systems. 2004quem mais cita →
- 10K Kuchnicka e D Maciejewski. Ventilator-associated lung injury. Anaesthesiology Intensive Therapy, 2013quem mais cita →
- 11Michigan Instruments. Mechanical Lung Simulator Model 5601i. 2015quem mais cita →
- 12S Mesić, R Babuska, H Hoogsteden e mais 1. Computer-controlled mechanical simulation of the artificially ventilated human respiratory system. IEEE Transactions on Biomedical Engineering, 2003quem mais cita →
- 13O Moerer. Weaning from mechanical ventilation: Which strategies are useful?. Anästhesiologie, Intensivmedizin, Notfallmedizin, Schmerztherapie, 2013quem mais cita →
- 14Q Mulqueeny, D Tassaux, L Vignaux e mais 4. Online estimation of respiratory mechanics in non-invasive pressure support ventilation: a bench model study. 2010quem mais cita →
- 15Oxymag. Magnamed. 2015
- 16E Saatçi e A Akan. Respiratory parameter estimation in linear lung models. 2008quem mais cita →
- 17PM Spieth, AR Carvalho, A Guldner e mais 9. Pressure support improves oxygenation and lung protection compared to pressure controlled ventilation and is further improved by random variation of pressure support. Critical Care Medicine, 2011quem mais cita →
- 18PM Spieth, AR Carvalho, P Pelosi e mais 10. Variable tidal volumes improve lung protective ventilation strategies in experimental lung injury. American Journal of Respiratory and Critical Care Medicine, 2009quem mais cita →
- 19PM Spieth, AR Carvalho, A Guldner e mais 4. Effects of different levels of pressure support variability in experimental lung injury. Anesthesiology, 2009quem mais cita →
- 20B Suki, SP Arold, R Mora e mais 2. Variable tidal volume ventilation improves lung mechanics and gas exchange in a rodent model of acute lung injury. American Journal of Respiratory and Critical Care Medicine, 2002quem mais cita →
Quem cita esta obra
nenhum artigo do acervo cita esta obra — ainda
É sua obra?
Receba um aviso quando um artigo novo citar este trabalho.
também serve para acompanhar o autor que você estuda