Dental Press Journal of Orthodontics2014

Insertion torque versus mechanical resistance of mini- implants inserted in different cortical thickness

Renata de Faria Santos, Antonio Carlos de Oliveira Ruellas, Daniel Jogaib Fernandes e Carlos Nelson Elias

em inglês

citado por 1 artigocita 16 obras, 3 no acervo

resumo

OBJECTIVE: This study aimed to measure insertion torque, tip mechanical resistance to fracture and transmucosal neck of mini-implants (MI) (Conexão Sistemas de PróteseT), as well as to analyze surface morphology. METHODS: Mechanical tests were carried out to measure the insertion torque of MIs in different cortical thicknesses, and tip mechanical resistance to fracture as well as transmucosal neck of MIs. Surface morphology was assessed by scanning electron microscopy (SEM) before and after the mechanical tests. RESULTS: Values of mechanical resistance to fracture (22.14 N.cm and 54.95 N.cm) were higher and statistically different (P < 0.05) from values of insertion torque for 1-mm (7.60 N.cm) and 2-mm (13.27 N.cm) cortical thicknesses. Insertion torque was statistically similar (P > 0.05) to torsional fracture in the tip of MI (22.14 N.cm) when 3 mm cortical thickness (16.11 N.cm) and dense bone (23.95 N.cm) were used. Torsional fracture of the transmucosal neck (54.95 N.cm) was higher and statistically different (P < 0.05) from insertion torsional strength in all tested situations. SEM analysis showed that the MIs had the same smooth surface when received from the manufacturer and after the mechanical tests were performed. Additionally, no significant marks resulting from the manufacturing process were observed. CONCLUSION: All mini-implants tested presented adequate surface morphology. The resistance of mini-implants to fracture safely allows placement in 1 and 2-mm cortical thickness. However, in 3-mm cortical thickness and dense bones, pre-drilling with a bur is recommended before insertion.

como citar · ABNT NBR 6023

SANTOS, Renata de Faria et al. Insertion torque versus mechanical resistance of mini- implants inserted in different cortical thickness. Dental Press Journal of Orthodontics, 2014. DOI: 10.1590/2176-9451.19.3.090-094.OAR.

Referências

16 obras citadas · 3 no acervo

  1. 1MA Papadopoulos e F Tarawneh. The use of miniscrew implants for temporary skeletal anchorage in orthodontics: a comprehensive review. Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 2007quem mais cita →
  2. 2M Barlow e K Kula. Factors influencing efficiency of sliding mechanics to close extraction space: a systematic review. Orthod Craniofac Res, 2008quem mais cita →
  3. 3PH Buschang, R Carrillo, B Ozenbaugh e mais 1. 2008 survey of AAO members on miniscrew usage. J Clin Orthod, 2008quem mais cita →
  4. 4CH Chen, CS Chang, CH Hsieh e mais 3. The use of microimplants in orthodontic anchorage. J Oral Maxillofac Surg, 2006quem mais cita →
  5. 5CH Moon, DG Lee, HS Lee e mais 2. Factors associated with the success rate of orthodontic miniscrews placed in the upper and lower posterior buccal region. Angle Orthod, 2008quem mais cita →
  6. 6HS Park, SH Jeong e OW Kwon. Factors affecting the clinical success of screw implants used as orthodontic anchorage. Am J Orthod Dentofacial Orthop, 2006quem mais cita →
  7. 7D Wiechmann, U Meyer e A Buchter. Success rate of mini- and micro-implants used for orthodontic anchorage: a prospective clinical study. Clin Oral Implants Res, 2007quem mais cita →
  8. 8R Reynders, L Ronchi e S Bipat. Mini-implants in orthodontics: a systematic review of the literature. Am J Orthod Dentofacial Orthop, 2009quem mais cita →
  9. 9ACO Ruellas, CT Mattos e CN Elias. Avaliação dos torques de inserção e remoção e da resistência mecânica de novos mini-implantes ortodônticos. Ortho Sci Orthod Sci Pract, 2012quem mais cita →
  10. 10CT Mattos, ACO Ruellas e CN Elias. Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study. Mat Res, 2010no acervo →
  11. 11CT Mattos, ACO Ruellas e EF Sant'Anna. Effect of autoclaving on the fracture torque of mini-implants used for orthodontic anchorage. J Orthod, 2011quem mais cita →
  12. 12Z Schwartz, CH Lohmann, G Blau e mais 3. Re-use of implant coverscrews changes their surface properties but not clinical outcome. Clin Oral Implants Res, 2000quem mais cita →
  13. 13TMN Araújo, MHA Nascimento, F Bezerra e mais 1. Ancoragem esquelética em Ortodontia com miniimplantes. Rev Dental Press Ortod Ortop Facial, 2006no acervo →
  14. 14MM Phiton, RL Santos, CN Elias e mais 2. In vitro evaluation of torsional strenght of orthodontic mini-implants. Braz J Oral Sci, 2008quem mais cita →
  15. 15GM Lima, MS Soares, SS Penha e mais 1. Comparison of the fracture torque of different Brazilian mini-implants. Braz Oral Res, 2011no acervo →
  16. 17A Bucheter, D Wiechmann, S Koerdt e mais 3. Load related implant reaction of mini-implants used for orthodontic anchorage. Clin Oral Implants Res, 2005quem mais cita →

Quem cita esta obra

1 artigo do acervo

quem fala desta obra

3 pessoas

É sua obra?

Receba um aviso quando um artigo novo citar este trabalho.

também serve para acompanhar o autor que você estuda
Acompanhar as citações