Patient-ventilator asynchrony

Marcelo Alcantara Holanda, Renata dos Santos Vasconcelos, Juliana Carvalho Ferreira e Bruno Valle Pinheiro

artigo de revisão · em inglês

citado por 3 artigoscita 60 obras, 1 no acervo

resumo

ABSTRACT Patient-v entilator asynchrony (PVA) is a mismatch between the patient, regarding time, flow, volume, or pressure demands of the patient respiratory system, and the ventilator, which supplies such demands, during mechanical ventilation (MV). It is a common phenomenon, with incidence rates ranging from 10% to 85%. PVA might be due to factors related to the patient, to the ventilator, or both. The most common PVA types are those related to triggering, such as ineffective effort, auto-triggering, and double triggering; those related to premature or delayed cycling; and those related to insufficient or excessive flow. Each of these types can be detected by visual inspection of volume, flow, and pressure waveforms on the mechanical ventilator display. Specific ventilatory strategies can be used in combination with clinical management, such as controlling patient pain, anxiety, fever, etc. Deep sedation should be avoided whenever possible. PVA has been associated with unwanted outcomes, such as discomfort, dyspnea, worsening of pulmonary gas exchange, increased work of breathing, diaphragmatic injury, sleep impairment, and increased use of sedation or neuromuscular blockade, as well as increases in the duration of MV, weaning time, and mortality. Proportional assist ventilation and neurally adjusted ventilatory assist are modalities of partial ventilatory support that reduce PVA and have shown promise. This article reviews the literature on the types and causes of PVA, as well as the methods used in its evaluation, its potential implications in the recovery process of critically ill patients, and strategies for its resolution.

como citar · ABNT NBR 6023

HOLANDA, Marcelo Alcantara et al. Patient-ventilator asynchrony. Jornal Brasileiro de Pneumologia, 2018. DOI: 10.1590/S1806-37562017000000185.

Referências

60 obras citadas · 1 no acervo

  1. 1H Wunsch, WT Linde-Zwirble, DC Angus e mais 3. The epidemiology of mechanical ventilation use in the United States. Crit Care Med, 2010quem mais cita →
  2. 2CS Barbas, AM Isola, AM Farias e mais 3. Brazilian recommendations of mechanical ventilation 2013 Part I. Rev Bras Ter Intensiva, 2014
  3. 3S Vasconcelos Rdos, LH Melo, RP Sales e mais 3. Effect of an automatic triggering and cycling system on comfort and patient-ventilator synchrony during pressure support ventilation. Respiration, 2013quem mais cita →
  4. 4DC Chao, DJ Scheinhorn e M Stearn-Hassenpflug. Patient-ventilator trigger asynchrony in prolonged mechanical ventilation. Chest, 1997quem mais cita →
  5. 5AW Thille, P Rodriguez, B Cabello e mais 2. Patientventilator asynchrony during assisted mechanical ventilation. Intensive Care Med, 2006quem mais cita →
  6. 6M de Wit, S Pedram, AM Best e mais 1. Observational study of patient-ventilator asynchrony and relationship to sedation level. J Crit Care, 2009quem mais cita →
  7. 7D Colombo, G Cammarota, M Alemani e mais 3. Efficacy of ventilator waveforms observation in detecting patient-ventilator asynchrony. Crit Care Med, 2011quem mais cita →
  8. 8C Alexopoulou, E Kondili, M Plataki e mais 1. Patientventilator synchrony and sleep quality with proportional assist and pressure support ventilation. Intensive Care Med, 2013quem mais cita →
  9. 9RD Branson. Patient-ventilator interaction the last 40 years. Respir Care, 2011quem mais cita →
  10. 10S Nava, C Bruschi, C Fracchia e mais 2. Patient-ventilator interaction and inspiratory effort during pressure support ventilation in patients with different pathologies. Eur Respir J, 1997quem mais cita →
  11. 11RS Vasconcelos, RP Sales, LHP Melo e mais 3. Influences of Duration of Inspiratory Effort, Respiratory Mechanics, and Ventilator Type on Asynchrony With Pressure Support and Proportional Assist Ventilation. Respir Care, 2017quem mais cita →
  12. 12JC Ferreira, DW Chipman, NS Hill e mais 1. Bilevel vs ICU ventilators providing noninvasive ventilation effect of system leaks: a COPD lung model comparison. Chest, 2009quem mais cita →
  13. 13A Jubran, WB Van de Graaff e MJ Tobin. Variability of patient-ventilator interaction with pressure support ventilation in patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med, 1995quem mais cita →
  14. 14MC Pohlman, KE McCallister, WD Schweickert e mais 3. Excessive tidal volume from breath stacking during lung-protective ventilation for acute lung injury. Crit Care Med, 2008quem mais cita →
  15. 15AS Slutsky e VM Ranieri. Ventilator-induced lung injury. N Engl J Med, 2013quem mais cita →
  16. 16S Nava, C Bruschi, F Rubini e mais 3. Respiratory response and inspiratory effort during pressure support ventilation in COPD patients. Intensive Care Med, 1995quem mais cita →
  17. 17JJ Marini, RM Rodriguez e V Lamb. The inspiratory workload of patientinitiated mechanical ventilation. Am Rev Respir Dis, 1986quem mais cita →
  18. 18NR MacIntyre, R McConnell, KC Cheng e mais 1. Patient-ventilator flow dyssynchrony flow-limited versus pressure-limited breaths. Crit Care Med, 1997quem mais cita →
  19. 19E Akoumianaki, A Lyazidi, N Rey e mais 3. Mechanical ventilation-induced reverse-triggered breaths A frequently unrecognized form of neuromechanical coupling. Chest, 2013quem mais cita →
  20. 20JR Beitler, SA Sands, SH Loring e mais 3. Quantifying unintended exposure to high tidal volumes from breath stacking dyssynchrony in ARDS the BREATHE criteria. Intensive Care Med, 2016quem mais cita →
1–20 de 60 obras citadas

Quem cita esta obra

3 artigos do acervo

ano a anomais citada em 2020 · 1 artigo
Mais citada em 2020, com 1 artigo.
  • 2020: citada em 1 artigo
  • 2021: citada em 1 artigo
  • 2026: citada em 1 artigo

quem fala desta obra

15 pessoas
ver as 15 pessoas e os artigos →

É sua obra?

Receba um aviso quando um artigo novo citar este trabalho.

também serve para acompanhar o autor que você estuda
Acompanhar as citações