Journal of the Brazilian Society of Mechanical Sciences and Engineering2010
Uncertainty calculation for hearing protector noise attenuation measurements by REAT method
Fabiano R. Lima, Samir Nagi Y. Gerges, Thiago Rodrigo L. Zmijevski, Darlam Fábio Bender e Rafael N. Cruz Gerges
em inglês
citado por 1 artigocita 10 obras
resumo
The objective of this paper is to quantify the uncertainty of hearing protector attenuation measurements and present the metrology study necessary for the accreditation of the Industrial Noise Laboratory (LARI) at The Federal University of Santa Catarina and Individual Protection Equipment laboratory (LAEPI) of NR consultancy Ltda. - Brazil for hearing protector noise attenuation procedures using the "Real Ear Attenuation at Threshold - REAT" method of the Brazilian National Institute of Metrology Standardization and Industrial Quality - INMETRO. A model for the calculation of measurement uncertainty was developed based on the document: "Guide to expression of uncertainty in measurement" published by the International Organization for Standardization, first edition, corrected and reprinted in 1995, Geneva, Switzerland. The uncertainty of each source of error was estimated and the overall uncertainty of the hearing protector noise attenuation measurement was calculated for each 1/1 octave band frequency and the results used in the single number (NRRsf - Noise Reduction Rating for subject fit) uncertainty calculation. It was concluded that the largest uncertainty is due to the determination of the subject hearing thresholds.
como citar · ABNT NBR 6023
LIMA, Fabiano R. et al. Uncertainty calculation for hearing protector noise attenuation measurements by REAT method. Journal of the Brazilian Society of Mechanical Sciences and Engineering, 2010. Disponível em: https://www.scielo.br/scielo.php?script=sci_arttext&pid=S1678-58782010000100005. Acesso em: 29 nov. 2026.
Referências
10 obras citadas
- 1ANSI. "ANSI-S 12.6-1997, American National Standards Method for measuring real-ear attenuation of hearing protectors". 1997quem mais cita →
- 2K Brinkmann e U Richter. "Variability and accuracy of sound attenuation measurements on hearing protectors". 1986quem mais cita →
- 3K Brinkmann. "Reapeatability and reproducibility of sound attenuation measurements on hearing protectors according to ISO 4869": PTB internal report. 1988quem mais cita →
- 4S.N.Y Gerges. "Ruído Fundamentos e Controle". 1992quem mais cita →
- 5C Guglielmone. "The uncertainty in the instrumentation for the measurement of noise". 2003quem mais cita →
- 6INMETRO, ABNT e SDM. "Guia para expressão da Incerteza de Medição": 1st Brazilian Edition of the "Guide to the expression of uncertainty in measurement". 1997quem mais cita →
- 7ISO. "ISO 4869-1: Acoustics - Hearing protectors - Part 1: subjective method for the measurement of sound attenuation". 1990quem mais cita →
- 8R.A Johnson. "Miller and Freund's Probability and Statistics for Engineers". 1994quem mais cita →
- 9F.R Lima. "Development and Metrologic Validation of a Hearing Protector Device Attenuation Measurement System by the Subjective Method". 2003quem mais cita →
- 10J.D Royster, E.H Berger, C.J Merry e mais 9. "Development of a new standard laboratory protocol for estimating the field attenuation of hearing protection devices: Part I - research of working group 11, accredited standards committee S12, noise". Journal of Acoustical Society of America, 1996quem mais cita →