Revista Latino-Americana de Enfermagem2019

Enzymatic detergent reuse in gastroscope processing: a potential source of microorganism transmission

Maria Letícia de Miranda Mati, Natália Rocha Guimarães, Paula Prazeres Magalhães, Luiz de Macêdo Farias e Adriana Cristina de Oliveira

em inglês

cita 37 obras, 1 no acervo

resumo

Objective: to evaluate the potential contamination of enzymatic detergent from its reuse and to identify the microbiological profile in the solution used to clean gastrointestinal endoscopic devices. Method: cross-sectional study based on microbiological analysis of 76 aliquots of 19 different enzymatic detergent solutions used to clean endoscopic devices. The aliquots were homogenized, subjected to Millipore® 0.45 µm membrane filtration and the presumptive identification of microorganisms was performed by biochemical-physiological methods according to previously established specific bacterial groups that are of clinical and epidemiological relevance. Results: the mean values, as well as the standard deviation and the median, of the enzymatic detergent microbial load increased as the solution was reused. There was a significant difference between the means of after first use and after fifth reuse. A total of 97 microorganisms were identified, with predominance of the coagulase-negative Staphylococcus, Pseudomonas spp., Klebsiella spp., Enterobacter spp. genus, and Escherichia coli species. Conclusion: the reuse of the enzymatic detergent solution is a risk to the safe processing of endoscopic devices, evidenced by its contamination with pathogenic potential microorganisms, since the enzymatic detergent has no bactericidal property and can contribute as an important source for outbreaks in patients under such procedures.

como citar · ABNT NBR 6023

MATI, Maria Letícia de Miranda et al. Enzymatic detergent reuse in gastroscope processing: a potential source of microorganism transmission. Revista Latino-Americana de Enfermagem, 2019. DOI: 10.1590/1518-8345.3101.3211.

Referências

37 obras citadas · 1 no acervo

  1. 1DB Nelson, AN Barkun, KP Block e mais 3. Transmission of infection by gastrointestinal endoscopy. Gastrointest Endosc, 2001quem mais cita →
  2. 2W Chapman. The principles of decontamination in gastrointestinal endoscopy. Br J Nurs, 2010quem mais cita →
  3. 3World Gastroenterology Organization e World Endoscopy Organization. Desinfecção de Endoscópios - um enfoque sensível aos recursos. 2011quem mais cita →
  4. 4American Society for Gastrointestinal Endoscopy. Technologies for monitoring the quality of endoscope reprocessing. Gastrointestinal Endoscopy, 2014quem mais cita →
  5. 5J Kovaleva. Infectious complications in gastrointestinal endoscopy and their prevention. Best Pract Res Clin Gastroenterol, 2016quem mais cita →
  6. 6A Srinivasan. Epidemiology and prevention of infections related to endoscopy. Curr Infect Dis Rep, 2003quem mais cita →
  7. 7N Kenters, EGW Huijskens, C Meier e mais 1. Infectious diseases linked to cross-contamination of flexible endoscopes. Endosc Int Open, 2015quem mais cita →
  8. 8Center for Disease Control and Prevention (CDC). New Dellhi-Metallo-b-lactamase-producing Escherichia coli Associated with endoscopic retrograde Cholangiopancreatography. MMWR Morb Mortal Wkly Rep, 2014quem mais cita →
  9. 9LF Muscarella. Risk of transmission of carbapenem resistant Enterobacteriaceae and related “superbugs” during gastrointestinal endoscopy. World J Gastrointest Endosc, 2014quem mais cita →
  10. 10Food and Drug Administration (FDA). Effective Reprocessing of Endoscopes used in Endoscopic Retrograde Cholangiopancreatography (ERCP) Procedures. 2015quem mais cita →
  11. 11LF Muscarella. ‘Updated’ Guidance for the Prevention of Transmission of Carbapenem-Resistant Enterobacteriaceae (‘CRE’) and Other Related Multidrug-Resistant ‘Superbugs’ during Gastrointestinal Endoscopy. Rev A Second Edition, 2016quem mais cita →
  12. 12IS Selhorst, MBC Bub e JBR Girondi. Protocolo de acolhimento e atenção para usuários submetidos a endoscopia digestiva alta e seus acompanhantes. Rev Bras Enferm, 2014no acervo →
  13. 13MA Bashaw. Guideline Implementation: Processing Flexible Endoscopes. Aorn J., 2016quem mais cita →
  14. 14U Beleihoff, H Biering, R Blum e mais 3. ESGE-ESGENA Reprocessing of flexible endoscopes and endoscopic accessories used in gastrointestinal endoscopy: Position Statement of the European Society of Gastrointestinal Endoscopy (ESGE) and European Society of Gastroenterology Nurses and Associates (ESGENA) - Update 2018. Endoscopy, 2018quem mais cita →
  15. 15U Beleihoff, CS Neumann, JF Rey e mais 3. ESGE-ESGENA Guideline for quality assurance in reprocessing: Microbiological surveillance testing in endoscopy. 2007quem mais cita →
  16. 16Ministério da Saúde (BR) e de 14 de novembro de 2012 Resolução da Diretoria Colegiada (RDC) nº 55. Dispõe sobre os detergentes enzimáticos de uso restrito em estabelecimentos de assistência à saúde com indicação para limpeza de dispositivos médicos e dá outras providencias. 2011quem mais cita →
  17. 17American Society for Gastrointestinal Endoscopy. Multisociety guideline on reprocessing flexible gastrointestinal endoscopes: 2011. Gastrointestinal Endoscopy, 2011quem mais cita →
  18. 18SS Evangelista, SG Santos, MAS Resende e mais 1. Analysis of microbial load on surgical instruments after clinical use and following manual and automated cleaning. Am J Infect Control, 2015quem mais cita →
  19. 19MJ Alfa. Current issues result in a paradigm shift in reprocessing medical and surgical instruments. Am J Infect Control, 2016quem mais cita →
  20. 20BT Petersen, CJ Cohen, RD Hambrick e mais 3. Multisociety guideline on reprocessing flexible GI endoscopes: 2016 update. Gastrointestinal Endoscopy, 2016quem mais cita →
1–20 de 37 obras citadas

Quem cita esta obra

nenhum artigo do acervo cita esta obra — ainda

É sua obra?

Receba um aviso quando um artigo novo citar este trabalho.

também serve para acompanhar o autor que você estuda
Acompanhar as citações