Revista Brasileira de Ortopedia2022

Is There Any Difference Between a Spherical Marker and a Simple Coin for Hip Replacement Digital Planning?

Giancarlo Cavalli Polesello, Thiago Tronco Salerno, Marcelo Cavalheiro de Queiroz, Walter Ricioli Junior, Nayra Deise Anjos Rabelo e Emerson Kyioshi Honda

em inglês

cita 23 obras, 1 no acervo

resumo

Abstract Objective To evaluate the accuracy and differences between 2 types of metallic markers, sphere, and coin, for radiographic calibration in the preoperative planning of hip arthroplasty. Methods Four spherical metallic markers and four coins, both 25 mm in diameter, were placed on the greater trochanter, pubic symphysis, between the thighs, and on the table of the exam, for radiographic examination of the hip in 33 patients with hip prosthesis. The prosthesis head was used for calibration and two examiners measured the markers' image diameters, and the results were analyzed statistically. Results In the greater trochanter, the sphere and the coin were not visualized in 19 radiographs (57.6%). Between the thighs, the coin marker was not visualized in 13 radiographs (39.4%). In the greater trochanter, the 25-mm accuracy of the coin and the sphere was, respectively, between 57.1 and 63.3% and between 64.3 and 92.9%. The coin between the thighs reached 25-mm accuracy in between 50 and 60% of cases. Over the exam table, the coin and sphere markers reached, respectively, the mean diameters of 22.91 mm and 23 mm, the lowest coefficient of variation, the lowest confidence interval, and the easiest positioning. There was statistical difference between the evaluations of the markers (coin vs. sphere) in all positions (p< 0.032), except for the exam table position (p= 0.083). Conclusions The coin between the thighs is the best marker for radiographic calibration in the preoperative planning of hip arthroplasty, and we suggest the use of another coin on the exam table for comparison, considering the 8% reduction in relation to its real size.

como citar · ABNT NBR 6023

POLESELLO, Giancarlo Cavalli et al. Is There Any Difference Between a Spherical Marker and a Simple Coin for Hip Replacement Digital Planning? Revista Brasileira de Ortopedia, 2022. DOI: 10.1055/S-0041-1724079.

Referências

23 obras citadas · 1 no acervo

  1. 1ME Muller. Total hip replacement: planning, technique and complications. In: Surgical management of degenerative arthritis of the lower limb, 1975quem mais cita →
  2. 2ME Müller. Lessons of 30 years of total hip arthroplasty. Clin Orthop Relat Res, 1992quem mais cita →
  3. 3JL Knight e RD Atwater. Preoperative planning for total hip arthroplasty. Quantitating its utility and precision. J Arthroplasty, 1992quem mais cita →
  4. 4V Hoikka, T Paavilainen, TS Lindholm e mais 2. Measurement and restoration of equality in length of the lower limbs in total hip replacement. Skeletal Radiol, 1987quem mais cita →
  5. 5S Eggli, M Pisan e ME Müller. The value of preoperative planning for total hip arthroplasty. J Bone Joint Surg Br, 1998quem mais cita →
  6. 6YM Baghdadi, AN Larson e RJ Sierra. Restoration of the hip center during THA performed for protrusio acetabuli is associated with better implant survival. Clin Orthop Relat Res, 2013quem mais cita →
  7. 7R Iorio, J Siegel, LM Specht e mais 3. A comparison of acetate vs digital templating for preoperative planning of total hip arthroplasty: is digital templating accurate and safe?. J Arthroplasty, 2009quem mais cita →
  8. 8S Wimsey, R Pickard e G Shaw. Accurate scaling of digital radiographs of the pelvis. A prospective trial of two methods. J Bone Joint Surg Br, 2006quem mais cita →
  9. 9M Franken, B Grimm e I Heyligers. A comparison of four systems for calibration when templating for total hip replacement with digital radiography. J Bone Joint Surg Br, 2010quem mais cita →
  10. 10MJ Archibeck, T Cummins e KR Tripuraneni. Inaccuracies in the Use of Magnification Markers in Digital Hip Radiographs. Clin Orthop Relat Res, 2016quem mais cita →
  11. 11KS Conn, MT Clarke e JP Hallett. A simple guide to determine the magnification of radiographs and to improve the accuracy of preoperative templating. J Bone Joint Surg Br, 2002quem mais cita →
  12. 12R Petretta, J Strelzow, NE Ohly e mais 2. Acetate templating on digital images is more accurate than computer-based templating for total hip arthroplasty. Clin Orthop Relat Res, 2015quem mais cita →
  13. 13P Gamble, J de Beer, D Petruccelli e mais 1. The accuracy of digital templating in uncemented total hip arthroplasty. J Arthroplasty, 2010quem mais cita →
  14. 14G Heinert, J Hendricks e MD Loeffler. Digital templating in hip replacement with and without radiological markers. J Bone Joint Surg Br, 2009quem mais cita →
  15. 15A González Della Valle, F Comba, N Taveras e mais 1. The utility and precision of analogue and digital preoperative planning for total hip arthroplasty. Int Orthop, 2008quem mais cita →
  16. 16B The, N Verdonschot, JR van Horn e mais 2. Digital versus analogue preoperative planning of total hip arthroplasties: a randomized clinical trial of 210 total hip arthroplasties. J Arthroplasty, 2007quem mais cita →
  17. 17CK Boese, P Lechler e L Rose. Calibration Markers for Digital Templating in Total Hip Arthroplasty. PLoS One, 2015quem mais cita →
  18. 18GC Polesello, TS Nakao e MC Queiroz. Proposta de Padronização do Estudo Radiográfico do Quadril e da Pelve. Rev Bras Ortop, 2011no acervo →
  19. 19CA Blake, J van der Merwe e JE Raubenheimer. A practical way to calibrate digital radiographs in hip arthroplasty. SA Orthop J, 2013quem mais cita →
  20. 20JL Fleiss. The design and analysis of clinical experiments. 1986quem mais cita →
1–20 de 23 obras citadas

Quem cita esta obra

nenhum artigo do acervo cita esta obra — ainda

É sua obra?

Receba um aviso quando um artigo novo citar este trabalho.

também serve para acompanhar o autor que você estuda
Acompanhar as citações