Química Nova2008
Determinação simultânea de creatinina e indicadores biológicos de exposição ao tolueno, estireno e xilenos em urina por cromatografia líquida de alta eficiência
Marina Venzon Antunes, Ana Luiza Marques Patuzzi e Rafael Linden
em português
resumo
A simple liquid chromatographic method for the simultaneous determination of creatinine, hippuric acid, mandelic acid, phenylglyoxylic acid and o, m and p-methylhippuric acids was developed and validated. Sample preparation was only dilution with water (1:10), followed by centrifugation. Analysis was performed in a reversed phase column (Lichrospher RP 8ec), 250 x 4.0 mm, with isocratic elution with phosphate buffer pH 2.3 and acetonitrile (90:10, v/v). The method presents adequate linearity, precision and accuracy and allows the simultaneous determination of the biomarkers of exposure to toluene, xylene and styrene together with creatinine, reducing cost and laboratory time.
como citar · ABNT NBR 6023
ANTUNES, Marina Venzon; PATUZZI, Ana Luiza Marques; LINDEN, Rafael. Determinação simultânea de creatinina e indicadores biológicos de exposição ao tolueno, estireno e xilenos em urina por cromatografia líquida de alta eficiência. Química Nova, 2008. DOI: 10.1590/S0100-40422008000700045.
Referências
2 obras citadas
Quem cita esta obra
nenhum artigo do acervo cita esta obra — ainda
É sua obra?
Receba um aviso quando um artigo novo citar este trabalho.
também serve para acompanhar o autor que você estuda