Revista da Sociedade Brasileira de Medicina Tropical2018
Is it possible to perform bacterial identification and antimicrobial susceptibility testing with a positive blood culture bottle for quick diagnosis of bloodstream infections?
Tamily Cristina Lemos, Laura Lúcia Cogo, Adriane Cheschin Maestri, Milena Hadad e Keite da Silva Nogueira
comunicação breve · em inglês
resumo
Abstract INTRODUCTION: Bloodstream infections can be fatal, and timely identification of the etiologic agent is important for treatment. METHODOLOGY: An alternative method, consisting of direct identification and susceptibility testing of blood culture bottles using the automated VITEK 2® system, was assessed. RESULTS: All 37 of the Gram-negative bacilli (GNB) identifications and 57.1% of the 28 Gram-positive cocci (GPC) identifications matched those obtained with standard methods. In susceptibility testing, the agreement was greater than 90%. CONCLUSIONS: This alternative methodology may assist in the early identification and susceptibility testing of GNB. Further research is necessary to develop appropriate methods for GPC.
como citar · ABNT NBR 6023
LEMOS, Tamily Cristina et al. Is it possible to perform bacterial identification and antimicrobial susceptibility testing with a positive blood culture bottle for quick diagnosis of bloodstream infections? Revista da Sociedade Brasileira de Medicina Tropical, 2018. DOI: 10.1590/0037-8682-0311-2017.
Referências
14 obras citadas · 1 no acervo
- 1CW Seymour, VX Liu, TJ Iwashyna e mais 3. Assessment of clinical criteria for sepsis for the Third International Consensus Definitions for Sepsis and Septic Shock. JAMA, 2016quem mais cita →
- 2T Lagu, MB Rothberg, MS Shieh e mais 3. Hospitalizations, costs, and outcomes of severe sepsis in the United States 2003 to 2007. Crit Care Med, 2012quem mais cita →
- 3A Kothari, M Morgan e DA Haake. Emerging technologies for rapid identification of bloodstream pathogens. Clin Infect Dis, 2014quem mais cita →
- 4A Lupetti, S Barnini, B Castagna e mais 2. Rapid identification and antimicrobial susceptibility testing of Gram-positive cocci in blood cultures by direct inoculation in to the BD Phoenix system. Clin Microbiol Infect, 2010quem mais cita →
- 5MJ Bruins, P Bloembergen, GJHM Ruijs e mais 1. Identification and susceptibility testing of Enterobacteriaceae and Pseudomonas aeruginosa by direct inoculation from positive BACTEC blood culture bottles into Vitek 2. J Clin Microbiol, 2004quem mais cita →
- 6LR Putnam, WJ Howard, MA Pfaller e mais 2. Accuracy of the Vitek system for antimicrobial susceptibility testing Enterobacteriaceae bloodstream infection isolates: use of “direct” inoculation from Bactec 9240 blood culture bottles. Diag Microbiol Infect Dis, 1997quem mais cita →
- 7MD Quesada, M Giménez, S Molinos e mais 3. Performance of VITEK-2 Compact and overnight Micro Scan panels for direct identification and susceptibility testing of Gram-negative bacilli from positive FAN BacT/ALERT blood culture bottles. Clin Microbiol Infect, 2010quem mais cita →
- 8M de Cueto, E Ceballos, L Martinez-Martinez e mais 2. Use of positive blood cultures for direct identification and susceptibility testing with the vitek 2 system. J Clin Microbiol, 2004quem mais cita →
- 9DT Leung, EM Davis, Q Qian e mais 1. First Report of prosthetic joint infection by Gemella sanguinis and associated “pseudosatelliting” phenomenon on culture. J Clin Microbiol, 2011quem mais cita →
- 10G Gherardi, S Angeletti, M Panitti e mais 3. Comparative evaluation of the Vitek-2 Compact and Phoenix systems for rapid identification and antibiotic susceptibility testing directly from blood culture of Gram-negative and Gram-positive isolates. Diagn Microbiol Infect Dis, 2012quem mais cita →
- 11TGS Paim, VV Cantarelli e PA d’ Azevedo. Performance of the Vitek2 system software version 5.03 in the bacterial identification and antimicrobial susceptibility test: evaluation study of clinical and reference strains of Gram-positive cocci. Rev Soc Bras Med Trop, 2014no acervo →
- 12T Spanu, M Sanguinetti, D Ciccaglione e mais 3. Use of the VITEK 2 System for rapid identification of clinical isolates of Staphylococci from bloodstream infections. J Clin Microbiol, 2003quem mais cita →
- 13KWT Ling, ZK Liu e AFB Cheng. Evaluation of the VITEK 2 System for rapid direct identification and susceptibility testing of Gram-negative bacilli from positive blood cultures. J Clin Microbiol, 2003quem mais cita →
- 14JH Jorgensen. Selection criteria for an antimicrobial susceptibility testing system. J Clin Microbiol, 1993quem mais cita →
Quem cita esta obra